Respiratory Humidifier T7
Respiratory Humidifier Taurus T7
Model : Taurus T7
- Taurus T7 Respiratory Humidifier is used to warm and humidify gases delivered to patients requiring mechanical ventilation, positive pressure breathing assistance, or other medical gases.
- The Servo Controlled Respiratory Heated Humidifier is designed to deliver optimally humidified gas to adult, neonatal and infant patients. This dual feedback system is designed to optimize humidity output while adjusting to changes in the environment.
Specifications:
Dimensions: | 135 mm x 165 mm x 155 mm (without chamber) |
Weight: | 2.9 Kg |
Display: | 3 Digit 14 mm 7 Segment LED |
Range: | 10 – 70 °C |
Accuracy: | ± 0.3 °C (in 20- 45 °C Temperature range) |
Supply Voltage: | 230 V |
Supply Frequency: | 50/60 Hz |
Supply Current: | 1.0 A max at 230 V |
Heater Plate Over-Temperature cut out: | 120± 6 °C |
Heater Plate: | 150W |
Heater Wire Supply : | 22V ± 5V~,2.75A Max |
Maximum Heater Wire Load: | 8 Ohm |
Mode Control Selection:
- Invasive mode: is used with patients whose upper airways have been bypassed by either a tracheotomy or endotracheal tube. Invasive mode is designed to deliver Optimal Humidity, for all therapies requiring gas delivered at 37°C, fully saturated. In this mode of operation, the humidifier attempts to deliver optimal humidity to the patient (37 °C, 100 % RH).
- Non-invasive: mode delivers comfortable levels of humidity, for face mask / headbox delivery . This mode is suitable only for patients whose natural humidification system (i.e. upper airways) has not been bypassed, but are receiving gas via a facemask. This aims to deliver gases at a comfortable level of humidity (31 °C, 32 mg/L) for face masks.
- Modes of operation: Heater Wire & Non Heater Wire
Performance:
Recommended Flow Range:
Invasive Mode: up to 60 L/min,
Non-invasive Mode: up to 120 L/min
Recommended ambient temperature range: 18 – 26 °C
Warmup Time: Less than 30 minutes
Alarms:
Airway / Chamber Probe(Sensor) Out Alarm, High Temperature Alarm, Low Temperature Alarm , Heater Wire & Temperature Probe alarm Alarms exceed 50 dBA @ 1m.
Standard Accessories:
- Temperature Probe
- Heater Wire Adapter
Heater Wire Mode:
Invasive Mode: Airway set point @40°C; Chamber set point @ 37°C
Non-Invasive Mode: Airway set point @34°C; Chamber set point @31°C
Non Heater Wire Mode:
Invasive Mode: Airway set point @37°C
Non-Invasive Mode: Airway set point @31°C
Standards:
IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6 , IEC 60601-1-8 ,EU CE